NeuPump 5B

  • Drug library with all the most used drugs
  • History record review and last configuration memory
  • Manual bolus and automatic bolus
  • Anti-bolus system to reduce significantly bolus after occlusion sudden release
  • Three kinds of mode: Rate Mode, Time Mode, Body Weight Mode
  • Accurate and sensitive pressure sensor detecting occlusion pressure accurately
  • Precision step motor, no pulsating wave even for low-rate injection
  • Selectable KVO rate according to concrete requirements from 0.1ml/h to 5ml/h
  • Accept all standard syringes of 5ml, 10ml, 20ml, 30ml and 50ml

Specifications

Applicable syringes

All brands of standard syringes: 5ml,10ml,20ml,30ml,50ml

Accuracy

±3%
Injection increment0.1ml
Occlusion value40kPa~160kPa
Injection rate5ml: 0.1 ml/h~100ml/h; 10ml: 0.1 ml/h~200ml/h
20ml: 0.1 ml/h~400ml/h; 30ml: 0.1 ml/h~600ml/h; 50ml: 0.1 ml/h~1500ml/h
Accumulated injection volume0.1ml~9999.9ml
Modes of operationSelectable: rate mode, time mode, body weight mode
Time modeVolume limit: 0.1~999.9ml Time: 1~2000 minutes
Body weight modeDrug limit:0.1~999.9mg
Solution limit:0.1~999.9ml
Weight: 01~300.0kg
Dose: 0.1~9999.9 mg/kg/h, ug/kg/min
Displayed informationRate, volume limit, accumulated injection volume, syringe size, pressure value, battery capacity, unit of medical dosage, bed No., AC power indicator, battery charge indicator, drug name, bolus rate
Alarm functionOVER, NEAR, Occlusion, Low battery, Syringe disengaged, No AC power supply, serious low battery
KVO rate0.1ml/h~5.0ml/h (selectable)
Bolus rate5ml: 100ml/h; 10ml: 200ml/h; 20ml: 400ml/h; 30ml: 400ml/h~600ml/h; 50ml: 400ml/h~1500ml/h
Weight1.8kg
Dimension310 X 125 X 115mm
Power supply
Fuse
AC100~220V, 50/60Hz
T2AL250AV
Maximum power consumption25W, can run for 4h after being recharged
BatteryRechargeable lithium polymer battery, 7.4V, 1600mAh
Battery rechargeWhen the pump is connected to the AC power supply, turn on the power
and the battery will automatically recharge (about 8~14 hours to recharge fully)
Electrical safetyComply with the stipulation of IEC60601-1: 1996
Electromagnetic compatibilityComply with the stipulation of IEC60601-1-2: 2006
Shell materialABS plastic
ClassificationII Class-BF
Operating conditionsEnvironment temperature: 5°C~40°C
air pressure: 80kPa~106kPa
relative humidity ≤ 80%
Storage conditionsEnvironment temperature: -20°C~50°C
air pressure: 50kPa~106kPa
relative humidity ≤ 95%